GE Healthcare
Revision 1
ERR SUB A003 - Gantry Error Messages
Code
Category
165/015
INHIBIT
165/016
INHIBIT
165/017
INHIBIT
165/018
INHIBIT
166/000
INHIBIT
166/001
INHIBIT
166/002
INHIBIT
166/003
INHIBIT
166/004
INHIBIT
166/005
INHIBIT
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Text in Log File or
on LCD
Internal software
error
Internal software
error
Internal software
error
Internal software
error
OPEN SOCKET
FAILED
CLOSE SOCKET
FAILED
ACCEPT CONNEC-
TION FAILED
TCP/IP PACKET
TRANSMISSION
FAILED
TCP/IP PACKET
RECEPTION
FAILED
TCP/IP PACKET
TOO LONG
Possible Cause/Recommended Action
Message list not defined.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Simulation Message description error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Simulation Message Flag description error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Message description error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Open socket failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Close socket failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The incoming connection was not accepted.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
A TCP/IP packet transmitted from the Gantry to
another node in the Senographe System failed
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry failed to receive a TCP/IP packet origi-
nating from another node in the Senographe Sys-
tem.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry received a TCP/IP packet originating
from another node in the Senographe System, but
rejected it because it was too long.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Page no. 924
Senographe DS
Job Card
ERR-SUB-A-003.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated