Revision 1
ERR SUB A003 - Gantry Error Messages
Code
Category
165/015
INHIBIT
165/016
INHIBIT
165/017
INHIBIT
165/018
INHIBIT
166/000
INHIBIT
166/001
INHIBIT
166/002
INHIBIT
166/003
INHIBIT
166/004
INHIBIT
166/005
INHIBIT
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Text in Log File or
on LCD
Internal software
error
Internal software
error
Internal software
error
Internal software
error
OPEN SOCKET
FAILED
CLOSE SOCKET
FAILED
ACCEPT CONNEC-
TION FAILED
TCP/IP PACKET
TRANSMISSION
FAILED
TCP/IP PACKET
RECEPTION
FAILED
TCP/IP PACKET
TOO LONG
Possible Cause/Recommended Action
Message list not defined.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Simulation Message description error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Simulation Message Flag description error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Message description error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Open socket failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Close socket failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The incoming connection was not accepted.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
A TCP/IP packet transmitted from the Gantry to
another node in the Senographe System failed
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry failed to receive a TCP/IP packet origi-
nating from another node in the Senographe Sys-
tem.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
The Gantry received a TCP/IP packet originating
from another node in the Senographe System, but
rejected it because it was too long.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.