GE Healthcare
Revision 1
ERR SUB A003 - Gantry Error Messages
Code
Category
170/006
DEBUG
ONLY
170/007
INHIBIT
170/008
INHIBIT
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Text in Log File or
on LCD
GRID NODE
In the case where the exposure time is between 20
ERROR
ms and 50 ms, the calculation is done using an expo-
sure time of 50 ms.
Ignore this error. No action is required.
Bucky Initialization
The Bucky failed to initialize with the Gantry.
Failed
1. Reboot the system from the X-ray Console.
2. If the problem persists, remove the carbon cover
3. If the problem persists, it could be linked to a
4. If the problem persists, change the Bucky and
Bucky Step Lost
When this error occurs intermittently it is due to a
bucky step loss.
When this error occurs frequently and there are grid
lines in the X-ray images it is due to a faulty or badly
adjusted Bucky connector cable (W310/W313 and
W311) located at the sides of the digital detector.
1. Reboot the system from the X-ray Console.
2. If the problem persists and there are grid lines in
3. If the problem persists and there are grid lines in
4. If the problem persists, change the Bucky and
Note: If this error occurs more than five times per
month change the Bucky.
Possible Cause/Recommended Action
and check that the Bucky can move.
mechanical issue, consequently add grease.
check the mechanical thickness.
the X-ray images, do the following with the Bucky
connector cable W310/W313:
a. Check that the cable connectors are fine.
Check that they are not disfigured or have
the plastic insulation covering them.
b. Check the cable continuity. You can do this
by looking for erratic variation in resistance
when gently wiggling the cable.
c. If the problem persists or this error occurred
during wiggling the cable, the cable is faulty.
the X-ray images, change the Bucky connector
cable W311.
a. Check that the cable connectors are fine.
Check that they are not disfigured or have
the plastic insulation covering them.
b. Check the cable continuity. You can do this
by looking for erratic variation in resistance
when gently wiggling the cable.
c. If the problem persists or this error occurred
during wiggling the cable, the cable is faulty.
check the mechanical thickness.
Page no. 930
Senographe DS
Job Card
D/R A255
page
1339
ELE A015
page 579
D/R A255
page
1339
D/R A255
page
1339
D/R A255
page
1339
ELE A015
page 579
ERR-SUB-A-003.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated