ERR SUB A003 - Gantry Error Messages
Code
Category
141/006
INHIBIT
141/007
INHIBIT
141/008
INHIBIT
141/009
INHIBIT
141/010
INHIBIT
141/011
INHIBIT
141/012
INHIBIT
141/013
INHIBIT
141/014
INHIBIT
Chapter 9
Text in Log File or
on LCD
Coll out of FOV
(front)
Coll out of FOV (left)
Coll out of FOV(rear)
Coll out of FOV
(right)
Coll filter type not
exist
Coll beyond phy limit
(front)
Coll beyond phy limit
(left)
Coll beyond phy limit
(rear)
Coll beyond phy limit
(right)
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
Invalid parameter for axis[i] (for i=1 - 5) = collimator
control board received a command for axis[i] that has
an invalid parameter. Either a wrong FOV value or a
wrong filter type.
Possible causes of error:
- Collimator has received bad data on its node
- Communication conflict/mismatch between
system and collimator
Actions to correct problem:
1. Check the cable connection (W409) between the
collimator and system, then restart system from
the X-ray Console.
2. If the problem persists, change the CPU board
(re-calibrate the collimator afterwards).
3. If the problem persists, contact your OLC.
axis[i](for i=1 - 4) beyond physical limit = the FOV is
beyond the blade capability.
1. Re-perform the collimator calibration. If the prob-
lem persists, identify the type of collimator cur-
rently in your system.
2. Replace the Collimator Set (refer to
R A280 - Collimator Set on page
3. Perform Collimator Format Calibration.
Page no. 902
Senographe DS
Job Card
D/R A190
page
1131
CAL A029
page
1683
D/R A280
page
Job Card D/
1369
1369).
CAL A029
page
1683
ERR-SUB-A-003.fm