Revision 1
13
AWS-IDC LINE TEST (EXTERNAL)
SUN VTS
Type of test:
For V1/V2 Control Stations from a command window, enter:
/opt/SUNWvts/bin/nettest –v -f -o dev=hme0,target=IDC
For V3 Control Stations from a command window, enter:
/opt/SUNWvts/bin/nettest –v -f -o dev=nge1,target=IDC
14
MOTHER BOARD SERIAL PORTS TESTS
14-1 Internal Test
SUN VTS
Type of test:
Note:
Motherboard serial port tests are only applicable to V1/V2 Control Stations because the V3 Control
Stations do not have onboard serial ports.
On V1/V2 Control Stations, from a command window, enter:
/opt/SUNWvts/bin/sptest –v –n -o L=I,T=a_b
JC-DIAG-A-001.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card DIAG A001 - AWS Sun-VTS Diagnostics
Page no. 985