GE Healthcare
Revision 1
cp error.log.back temp
h. If you want the older log files, copy them by entering:
cp trace.SenoServer.log.xxxxxxxxxxxxxx
i.
If you want the older archived log files, copy them by entering:
cp oldfiles/trace.SenoServer.log.xxxxxxxxx.Z
3. Change to the temporary (/export/home/sdc/senovision/logfiles/temp) directory, by entering:
cd temp
4. Compress all of the files in the temporary (/export/home/sdc/senovision/logfiles/temp) directory, by
entering:
compress *
5. If any of the resulting files exceed the size of a floppy disk, they can be divided into smaller parts by
using the split command. For example, to split the file ExamEvents.log.Z into parts smaller than
1.3 Mbyte, enter the following command:
split -b 1300 k ExamEvents.log.Z
This generates two part files: ExamEvents.log.Z.01 and ExamEvents.log.Z.02
Each of the part files can then be copied to floppy as for the original files.
5-4-3
Copy Log Files to Floppy Disk
Copy the required files to a floppy disk as follows:
1. Insert a formatted disk in the floppy drive.
2. Log in as root as follows:
a. At the prompt, enter:
su - root
b. When requested to specify the root password, enter:
operator
3. Mount the floppy drive device as /floppy by entering:
mount -F pcfs /dev/fd0 /floppy
4. Change to the temporary (/export/home/sdc/senovision/logfiles/temp) directory where the com-
pressed log files exist, by entering:
cd /export/home/sdc/senovision/logfiles/temp
5. For each required log file, enter the following command to copy the appropriate file to floppy disk:
cp <logfile>.log.Z /floppy
For each file, replace the string <logfile> by the appropriate file name.
To copy non-compressed files, omit the characters .Z from the command.
5-4-4
Completion
After copying all required log files, perform the following steps to eject and unmount the floppy disk, and
to remove the compressed log files you created.
1. Return to the root directory by entering:
cd
2. Eject the floppy disk by entering:
eject
3. Unmount the floppy device by entering:
umount /floppy
4. Close the xterm window by entering:
JC-DIAG-A-005.fm
Service Information and Procedures Class A 2385072-16-8EN
Page no. 1003
Job Card DIAG A005 - Error Log Retrieval
Senographe DS
Chapter 9
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated