GE Healthcare
Revision 1
ERR SUB A001 - Generator Error Messages
Code
Message text
027/005 E04 Arm fatal failure A Gantry error was signaled during the
027/006 E04 Arm fatal failure A Gantry error was signaled during the
031/001
031/002
041/001
041/002
041/003
041/004
041/005
041/006
050/043
052/001 E04 Arm fatal failure No Gantry acknowledgment to change
056/001
056/002
056/003
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Description
exposure. Occurs at the end of expo-
sure when the Generator requests the
Gantry to stop the grid motion. If the
Gantry does not answer within 3 sec of
the request, this error occurs.
exposure.
Generator: delay task software error:
unknown command
Generator: delay task software error:
unknown task status
Gene thermal protection software:
VRTX error
Gene thermal protection software:
VRTX error
Gene thermal protection software:
VRTX error
Gene thermal protection software:
VRTX error
Gene thermal protection software:
unknown anode starter status
Gene thermal protection software:
VRTX error
An unexpected request was received in
SysCntrl.
a filter request.
Focus software error: ACQ_NTR or
ACQ_OVR received or unknown reply
Focus software error: unknown com-
mand
Focus software error: unknown task
status
Page no. 838
Senographe DS
Recommended actions
1. Reset Senographe
2. Check communications between
Gantry and Generator.
1. Reset Senographe
2. Check communications between
Gantry and Generator.
Report the problem to engineering.
Check the error logs.
If there are any collimator reported
errors change the impacted collima-
tor-related FRUs.
If there are no collimator reported
errors change the other impacted sys-
tem-related FRUs.
ERR-SUB-A-001.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.