GE Senographe DS Manual page 294

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Central Listing
4-5. Bucky Board PL302
Board-Bucky PL302; see component index page 157
DS2
DS3
DS6
J3
LED
DS1
DS2
Data CAN reception
DS3
RT2 Synchro Grid
DS4
DS5
DS6
Microcontroller Debug
DS7
Microcontroller Heartbeat
DS8
Microcontroller Debug
DS9
DS10
DS11
Optical Fork Behavior
Chapter 3
J1
DS8
DS7
DS9
DS10
J4
Function
48 V
Not present
Not present
5 V
3.3 V logic
Service Information and Procedures Class A 2385072-16-8EN
.
J2
Color
Green
ON when the 48 V DC supply voltage is OK
OFF when the 48 V DC supply voltage is defec-
tive
Yellow
ON when CAN data (or command) received.
OFF when no CAN data (or command) received.
Yellow
n/a
n/a
n/a
n/a
n/a
Green
n/a
Green
Blinks once every second when the Bucky board
is okay
Blinks more than once a second or is constantly
ON or constantly OFF when the bucky board is
faulty
Green
n/a
Green
ON when the 5 V supply voltage is OK
OFF when the 5 V supply voltage is defective
Green
ON when the 3.3 V supply voltage is OK
OFF when the 3.3 V supply voltage is defective
Green
ON when the optical fork on the Bucky is
masked (i.e. the Bucky grid is initialized)
OFF when the optical fork on the Bucky is not
masked (i.e. the Bucky grid is no initialized)
Page no. 294
Senographe DS
J5
DS11
Behavior
Indicators and Switches.fm