GE Senographe DS Manual page 863

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Code
Category
102/047
DEBUG_O
NLY
102/048
DEBUG_O
NLY
102/049
DEBUG_O
NLY
102/050
HALT
102/051
HALT
102/052
DEBUG_O
NLY
102/053
DEBUG_O
NLY
103/001 ADVISORY
ERR-SUB-A-003.fm
Text in Log File or
on LCD
LIFT INVALID INI-
TIALIZATION
ERROR
LIFT INVALID
FORCE SENSOR
DEFECT ERROR
LIFT INVALID
FORCE SENSOR
RANGE ERROR
LIFT HBRIDGE
VOLTAGE ERROR
LIFT UNEXPECTED
EVENT ERROR
LIFT INVALID GO
ROTATION FAIL-
URE ERROR
LIFT INVALID GO
ANGULATION FAIL-
URE ERROR
LCD: A06 Lift end
position is reached
Service Information and Procedures Class A 2385072-16-8EN
Possible Cause/Recommended Action
There has been an initialization error. The Lift node
normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
There has been a force sensor defect. The Lift node
normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
There has been a force sensor range error. The Lift
node normally does not report this error in any case.
Ignore this error. If the problem persists, contact your
OLC.
There has been an H-Bridge voltage error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
There has been a software error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
On every "StartMotion" command, the IPMS checks
whether the software stop limit for the specified
direction is not already reached. If the limit is
reached, the "StartMotion" command is ignored and
the following error is returned:
Lift
"UpperLimitAlreadyReached" if the lift upper stop
limit is active
"LowerLimitAlreadyReached" if the lift lower stop
limit is active
Operator error. No service action is required.
Page no. 863
ERR SUB A003 - Gantry Error Messages
Senographe DS
Job Card
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Subscribe to Our Youtube Channel

Table of Contents

Save PDF