4-3. PDU Board PL101
Board-PDU PL101; see component index page 158
Fuse
F1
F3
F4
F5
F6
Jumper
PRGM
Used to program the PDU board during manufacturing. Closed to program
J22
Used to turn on/off 48 V when the PDU board is in programming mode. Bot-
tom two pins closed turn 48 V off. Top two pins closed turn 48 V on.
Open enables normal PDU board operation, where the 48 V is controlled by
LED
DS1
DS2
DS3
Minus sign used for error codes
DS4
Minus sign used for error codes
Chapter 3
J4
J2
DS2
DS1
F5
DS26
F6
J1
DS21‐20‐19‐17‐16
Specification
250 Vac, 6.25 A, Delay
250 Vac, 0.5A, Delay
250 Vac, 2A, Delay
250 Vac, 125 mA, Delay
the PDU board. Open for normal PDU board operation.
the Senographe system.
Function
DMR On
48 V
Service Information and Procedures Class A 2385072-16-8EN
.
J5
J7
F3
F4
DS24
S4
PRGM
DS25
DS15, 18
J3
46-170021P60
46-170021P31
46-170021P30
5344386-2
Function
Color
Green
ON when the Generator is ON
OFF when the Generator is OFF
Green
ON when 48 V DC supply voltage is OK
OFF when 48 V DC supply voltage is defective
Red
Sometimes ON when reporting certain error codes.
Red
Page no. 290
J6
S3
S2
S1
F1
DS5
DS6
DS7
DS8
J8
J9
J22
DS14‐13‐11‐10‐12‐9
J10
J11
PN
Behavior
Senographe DS
Protects
Position
This jumper must
always be open.
This jumper must
always be open.