R&S
ZNB/ZNBT
Defaults are ±0.2 dB for magnitude and ±2° for phase deviations.
Remote command:
CALCulate:CALValidate:RESPonse:MAGNitude:LOWer
CALCulate:CALValidate:RESPonse:MAGNitude:UPPer
CALCulate:CALValidate:RESPonse:PHASe:LOWer
CALCulate:CALValidate:RESPonse:PHASe:UPPer
Preferences
Allows you to configure the behavior of the validation diagrams and traces.
Unchecking "Auto Scale" prevents the traces from scaling as the measurement and/or
standard changes. Unchecking "Average On" turns off averaging.
Basic Mode
Switches to
5.11.2 Cal Devices tab
Provides access to all functions for calibration kit management and cal unit characteri-
zation.
Background information
Refer to the following sections:
●
Chapter 4.5.2, "Calibration standards and calibration
●
Chapter 4.5.5.3, "Characterization of calibration
The buttons in the "Cal Devices" tab open the following dialogs:
●
"Cal Connector Types...": See
on page 512
●
"Cal Kits...": See
User Manual 1173.9163.02 ─ 62
Basic mode
Chapter 5.11.2.2, "Calibration Kits
kits",
units",
Chapter 5.11.2.1, "Cal Connector Types
dialog",
GUI reference
Cal softtool
on page 174
on page 185
dialog",
on page 514
511