3. "SMARTerCal":
a) Acquire measurement data for all required standards for the selected system
b) Apply the calculated error terms to the active channel.
c) Acquire the power calibration data at the selected source port.
d) Use the source power calibration data to correct the absolute receiver powers
Background and related information
Refer to
Use the
Chapter 4.5.7.3, "Combining SMARTerCal with scalar power
on page 200).
When you apply the acquired "SMARTerCal", the active calibration is replaced and
discarded.
To persist any kind of calibration, you can transfer it to the calibration "Pool" using
the
The system error correction data determined in a calibration procedure is stored in
the analyzer. You can read these correction data using the remote control com-
mand [SENSe<Ch>:]CORRection:CDATa.
If external switch matrices are involved, use
SMATrix:CDATa
Step 1: Ports
Selects the calibrated analyzer ports, the cal unit characterization, the calibration type
and the power meter connection.
Background information
Refer to the following sections:
Chapter 4.5.7.2, "Calibration
Chapter 4.5.5.3, "Characterization of calibration
User Manual 1173.9163.02 ─ 62
error calibration type.
at all calibrated ports.
Chapter 4.5.7,
"SMARTerCal",
Power Cal wizard
Calibration Manager
dialog.
instead.
on page 199 for background information
to perform an additional source flatness calibration (see
[SENSe<Ch>:]CORRection:
types",
on page 200
units",
GUI reference
Cal softtool
calibration",
on page 185
483