R&S ZNB Series User Manual page 378

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Controls on the Define Limit Lines dialog
The "Define Limit Lines" dialog contains a table to edit the individual segments of the
limit lines. The buttons below the table extend or shorten the segment list.
Creating limit lines with minimum effort
Choose one of the following methods to create and handle limit lines efficiently:
To define limit lines with only a few segments, select "Add" and edit each segment
in the
Select a data or memory trace as a limit line ("Import Trace...") or import a trace
stored in a file ("Import File...").
Save your limit lines to a file so you can reuse or modify them later sessions ("Save
Limit Line...", "Recall Limit Line...").
Segment List
Defines the individual limit line segments.
The table contains an automatically assigned current number for each segment plus
the following editable columns:
"Type" indicates whether the segment belongs to an "Upper" or a "Lower" limit line,
or if the limit check at the segment is switched "Off". Switching off the limit check
does not delete the segment but changes its screen color.
"Start Stimulus" is the stimulus (x-axis) value of the first point of the segment.
"Stop Stimulus" is the stimulus (x-axis) value of the last point of the segment.
"Response"
Tap the "Response" field to open the
configuration possibilities.
"Interpolation" determines whether the limit line segment is interpolated linearly or
logarithmically. The latter is possibly more suitable for logarithmic sweeps.
Remote command:
CALCulate<Chn>:LIMit:SEGMent:COUNt?
CALCulate<Chn>:LIMit:SEGMent<Seg>:TYPE
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STARt
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STOP
User Manual 1173.9163.02 ─ 62
Segment List
individually.
For a limit line segment that is defined as linear functions of the stimulus,
"Response" displays the y-axis values at the start and stop stimulus (x-axis)
values.
For formula-defined line segments, it displays the formula.
Response dialog
that offers the full range of
GUI reference
Lines softtool
378

Advertisement

Table of Contents
loading

Table of Contents