a) Acquire error correction data for the required ports and port pairs, and the
b) Acquire source power calibration data at the "Power Port".
c) Finally, decide whether to apply the resulting calibration.
Background and related information
Refer to
See
the "MultiCal" variant of the wizard.
Use the
Chapter 4.5.7.3, "Combining SMARTerCal with scalar power
on page 200).
If the active channel is already calibrated, the wizard attempts to load the calibra-
tion. If the calibration setup is unchanged and sweep data are available, the exist-
ing system error correction can be optimized without repeating the measurement of
all standards.
When you apply the acquired SMARTerCal, the active calibration is replaced and
discarded.
To persist any kind of calibration, you can transfer it to the cal "Pool" using the
bration Manager
The active system error correction data can be read (and modified) using the
remote control command [SENSe<Ch>:]CORRection:CDATa.
If external switch matrices are involved, use
SMATrix:CDATa
Step 1: Ports and Type
Selects the calibrated analyzer ports, the calibration type and specifies the power
meter connection.
Background information
Refer to
User Manual 1173.9163.02 ─ 62
required standards.
Chapter 4.5.7,
"SMARTerCal",
Chapter 5.11.1.11, "SMARTerCal wizard (manual,
Power Cal wizard
dialog.
instead.
Chapter 4.5.7.2, "Calibration
on page 199 for background information.
to perform an additional source flatness calibration (see
[SENSe<Ch>:]CORRection:
types",
on page 200.
GUI reference
Cal softtool
MultiCal)",
on page 505 for
calibration",
Cali-
497