Tracking
Enables or disables tracking for "Multiple Peak" search. If enabled, a new multiple
peak search is performed for each sweep (creating/deleting markers as needed).
Define an
val.
Remote command:
CALCulate<Chn>:MARKer<Mk>:SEARch:TRACking
Marker Config...
Opens the
All Markers Off
See
5.7.4.2
Multiple Marker Config dialog
The "Multiple Marker Config" dialog allows you to configure the multiple peak searches
for the active trace.
Access: Trace – [Marker] > "Multiple Peak" > "Marker Config..."
Search Config
Same as selecting
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MMAXimum | MMINimum
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:EXECute MMAXimum |
MMINimum
Threshold Settings
Defines a threshold for multiple peak searches and activates it.
If "Active", only maxima above and minima below the configured "Threshold" are con-
sidered.
The unit is derived from the active trace format and cannot be modified. Changing the
trace format resets the "Threshold" to a format-specific default value.
Remote command:
CALCulate<Chn>:MARKer:MPEak:THReshold
CALCulate<Chn>:MARKer:MPEak:THReshold:STATe