Switching on the display is ideal for program test purposes but tends to slow down the
measurement. Therefore, it is recommended to switch off the display in real measure-
ment applications where a tested program script is to be executed repeatedly.
The analyzer provides a third display option where the measurement screen is only
updated when triggered by the remote control command SYSTem:DISPlay:UPDate
ONCE.
The instrument remains in the remote state until it is reset to the manual state via the
GUI or via remote control (see
on page 810). You can also lock the remote (touch) screen using SYSTem:TSLock
SCReen.
A tooltip across the bottom of the remote screen indicates a remote command error.
You can switch off this tooltip using SYSTem:ERRor:DISPlay:REMote OFF.
SCPI commands:
@REM
SYSTem:DISPlay:UPDate
SYSTem:TSLock
SYSTem:ERRor:DISPlay[:REMote]
6.1.4.1
Setting the device address
The GPIB address (primary address) of the instrument is factory-set to 20. It can be
changed manually in the System – [Setup] > "Remote Settings" tab or via remote con-
User Manual 1173.9163.02 ─ 62
Chapter 6.1.4.2, "Returning to manual
Remote control
Introduction to remote control
operation",
809