Selects the test ports to be calibrated.
It is possible to select any combination of two or more test ports. If you are only inter-
ested in a single port p, perform a two-port SMARTerCal for a port pair including p (as
Power Port).
Note: Calibration and port de-/activation.
The analyzer fimware automatically activates/deactivates ports during/after a (success-
ful) calibration:
Calibrated ports that were previously disabled, are automatically enabled as single-
ended logical ports.
An uncalibrated port that is not used by a measurement (i.e. the port is not required
by any trace of the related channel) is disabled.
Remote command:
The port parameters in many calibration commands define ports to be calibrated.
Type
Selects the calibration type to be performed. For background information, see
ter 4.5.7.2, "Calibration
SMARTerCal is based on a "One Path Two Ports" or a full n-port calibration (n > 1):
"Adapter Removal" is a 2-port calibration, all other calibration types can be used for
any combination of 2 or more ports.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID
Power Port
Selects the port for the scalar source power calibration.
Source leveling
If checked, the "SMARTerCal" wizard opens the
There you can perform a complementary
Cal"ibrated" ports, and additional scalar power calibrations at other ports.
Select All/Deselect All
Selects/deselects all ports.
These buttons are only shown if more than 4 ports are available.
Next
Proceeds to
(and a warning is displayed) if fewer than two ports are selected.
User Manual 1173.9163.02 ─ 62
types",
on page 200.
"Step 2: Connectors and Cal Kits"
"Configured Devices"
on page 782.
Power Cal wizard
source flatness calibration
on page 500. This button is inactive
GUI reference
Cal softtool
Chap-
on completion.
at the ""SMARTer-