R&S
ZNB/ZNBT
The available settings comprise the sweep type (with related parameters), the trigger
conditions, and the periodicity of the measurement.
Access: Channel – [Sweep] hardkey
Background information
Refer to the following sections:
Chapter 4.1.4, "Sweep
Chapter 4.1.4.4, "Stimulus and sweep
5.10.1 Sweep Params tab
Allows you to define the scope and timing of the measurement in the active channel.
Segmented sweeps
In "Segmented" linear frequency sweeps, the sweep parameters can be set independ-
ently for each segment. See
System error correction
In general, the system error correction is no longer valid after a change of the sweep
parameters. The status of the calibration is shown in the trace list. If the number of
points is changed, the analyzer interpolates the correction data. The calibration label
"Cal Int" is displayed. See also
Number of Points
Sets the total number of measurement points per sweep. The minimum number of
points is 1 (measurement at a single frequency/power/time value), the maximum is
100,001.
User Manual 1173.9163.02 ─ 62
control",
on page 90
types",
Chapter 5.10.2.2, "Define Segments
Chapter 4.5.4, "Calibration
GUI reference
Sweep Softtool
on page 96
dialog",
labels",
on page 181.
on page 431.
423