R&S ZNB Series User Manual page 482

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
left = "Ref. Receiver"
right = "Power Meter"
Start Cal Sweep
Start the calibration sweeps for the selected port and power calibration settings and
close the dialog. The calibration is performed as described in
on page 195.
Use the
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire]
5.11.1.8
SMARTerCal wizard (Cal Unit)
The "SMARTerCal (Cal Unit)" wizard guides you through the setup and execution of an
automatic "SMARTerCal".
Access: Channel – [Cal] > "Start Cal" > "SMARTerCal" – "Start... (Cal Unit)"
The guided "SMARTerCal" is analogous to a regular automatic calibration with an addi-
tional power calibration sweep at a single analyzer port. It consists of the following
steps:
1. "Ports": Select the calibrated ports, the cal unit characterization and the calibra-
tion type.
2. "Connections":
a) Define the port assignment between the R&S ZNB/ZNBT and the calibration
b) Select one of the calibrated ports for power calibration.
User Manual 1173.9163.02 ─ 62
Pwr Cal Method
Pwr Cal Method
Pwr Cal Settings
tab"Pwr Cal Settings" to modify the calibration procedure.
unit.
GUI reference
Cal softtool
"Calibration procedure"
482