R&S ZNB Series User Manual page 125

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
The following table gives an overview of recommended display formats.
Complex dimensionless quan-
tities:
S-parameters and ratios
Lin Mag
ON
dB Mag
ON (default)
Phase
ON
Real
ON
Imag
ON
Unwrapped Phase
ON
Smith
ON (reflection coefficients S
Polar
ON
Inverted Smith
ON (reflection coefficients S
SWR
ON (reflection coefficients S
Delay
ON (transmission coefficients S
The default formats are activated automatically when the measured quantity is
changed.
4.3 Measurement results
This section gives an overview of the measurement results of the network analyzer and
the meaning of the different measured quantities. All quantities can be selected in the
"Meas" softtool (function key Trace – [Meas]).
4.3.1 S-parameters
S-parameters are the basic measured quantities of a network analyzer. They describe
how the DUT modifies a signal that is transmitted or reflected in forward or reverse
direction. For a 2-port measurement, the signal flow is as follows.
User Manual 1173.9163.02 ─ 62
normalized impedance or admittance) if they are used for transmission S-parame-
ters, ratios and other quantities.
For complex "Impedances", "Admittances", "Z-parameters", and "Y-parameters"
generally a Cartesian format or the polar format is suitable.
For the real valued
Stability factors
sian formats "Lin Mag" or "Real" should be used. In complex formats, real numbers
represent complex numbers with zero imaginary part.
)
ii
)
ii
)
ii
)
ij
and DC measurement results, one of the Carte-
Complex quantities with dimensions:
Wave quantities, Z-parameters, Y-
parameters, impedances, admittan-
ces
ON (default for Z-parameters, Y-param-
eters, impedances, admittances)
ON (default for wave quantities)
ON
ON
ON
ON
Concepts and features
Measurement results
Real quantities:
Stability Factors, DC 1 ... 4,
PAE
ON (default)
ON
125