ZNB/ZNBT
The limit line colors are defined in the "Define User Color Scheme" dialog (see
ter 5.17.3.2, "Define User Color Scheme
between various options:
Assign the same color to traces and associated limit lines.
Assign different colors to limit line segments with disabled limit check.
Note: Limit line display and limit check are independent of each other: Hiding the limit
line does not switch off the limit check.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:DISPlay[:STATe]
Limit Check
Switches the limit check of the active trace on or off. When the limit check is switched
on, a movable "PASS" or "FAIL" message is displayed in the diagram.
Limit violations are marked with squares, whose color is defined in the
Color Scheme dialog
A
TTL signal
User Manual 1173.9163.02 ─ 62
("Limit Fail Trace Color").
indicating pass or fail can be generated in addition.