ZNB/ZNBT
Remote command:
The port set number corresponds to the <ListId> numeric suffix in the port set de-/
embedding commands; see e.g.
DEEMbedding<ListId>[:STATe].
All Deembedding Activated
Deactivates or activates all deembeddings (single-ended, balanced, port set, ground
loop) for the active channel, if any.
This control is only visible if
Network
Depending on the size m of the selected port set, the graphical list contains all availa-
ble 2m-port deembedding/embedding networks.
Note: For port pairs (m=2) the deembedding/embedding network can be defined either
via lumped element models (in combination with s2p Touchstone files) or via a s4p
Touchstone file (see
m≥3, there are no predefined lumped element models available; the deembedding/
embedding network has to be defined via an s<2m>p Touchstone file.
The
set. The "D1" and "D2" networks are defined by imported S-parameter data; see
D2.
For 4-port networks that are (partly) defined by lumped elements, the lumped element
parameters are displayed below the graphical network list. See
cuit models for 4-port
Tip:
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol. The lumped elements and S-parameter networks ("D1", "D2") are
numbered from top to bottom.
Use the
in "Capacitor in parallel with resistor" circuit blocks to conductances and vice versa.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:C<1|2|3>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:L<1|2|3>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:R<1|2|3>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:G<1|2|3>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:
PARameters:C<1|2|3>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:
PARameters:L<1|2|3>
User Manual 1173.9163.02 ─ 62
Type
Chapter 4.6.2.5, "Port pair
symbol selects "no network" and disables de-/embedding for the selected port
networks",
Conductance in Embedding Networks
CALCulate<Ch>:TRANsform:VNETworks:PPAir:
= "Deembedding".
de-/embedding",
on page 212.
switch to change from resistances
GUI reference
Offset Embed softtool
on page 213). For
D1,
Chapter 4.6.2.4, "Cir-
642