The name of a characterization file inside the zip container must indicate
the standard type (offset short, mismatch, 20 dB Att, 40/50 dB Att, airline, mis-
match airline, att, attenuator, stepped thru)
and, if applicable, its gender (f, m, fm, ff, mf, mm).
Imported verification kits with uncertainties are preserved and added to the
Kit
selection combo-box.
Step 2: METAS Verification
Step 2 allows you to perform the verification.
By selecting a button representing a verification standard, the required measurements
are performed. The raw measurement data are handed over to METAS VNA Tools that
returns the calibrated results and uncertainties and the pass/fail states of the respec-
tive verifications. The latter are indicated on the trace fields to the right of the verifica-
tion standard buttons (red = failed).
The verification standard buttons indicate the verification states of the respective verifi-
cation standards ("V" = not verified yet, "✓" = passed, "X" = failed). Use the
button to toggle a condensed view focusing on these verification states.
Collapse
In collapsed state, the "METAS Verification" dock widget presents a condensed view,
focusing on the verification results for the available verification standards.
Save Verification Data
Same as
the METAS verification configuration, and the verification measurement data obtained
so far.
User Manual 1173.9163.02 ─ 62
Export Data to METAS VNA
Tools. The exported ZIP file now also contains