R&S
ZNB/ZNBT
Figure 5-57: Measure Fixture dialog: Direct Compensation (and Prompt for Each Port)
"Direct Compensation" data files are standard trace files, containing reflection parame-
ter traces for the related port and standard:
Table 5-5: Direct Compensation data
Prompt for Each Port
disabled
enabled
Tip: Remote control provides additional flexibility. You can:
●
Measure the same ports repeatedly without changing the standards and attribute
the results to different channels.
●
Calculate the compensation data for different ports, using mixed Open and Short
standards.
Refer to the program example for
FIXTure[:ACQuire].
Remote command:
[SENSe<Ch>:]CORRection:COLLect:FIXTure:STARt
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
[SENSe<Ch>:]CORRection:COLLect:FIXTure:SAVE
[SENSe<Ch>:]CORRection:COLLect:FIXTure:EXPort
[SENSe<Ch>:]CORRection:COLLect:FIXTure:IMPort
5.13.3 One Way Loss tab
Defines loss parameters for each physical port.
Use the complementary dock widget to activate or deactivate length/loss compensation
for selected ports (see
User Manual 1173.9163.02 ─ 62
File Type
csv
s1p
[SENSe<Ch>:]CORRection:COLLect:
Chapter 5.13.1.2, "Offset
Offset Embed softtool
Description
One csv trace file per standard,
containing reflection traces for all
selected ports; see
ter 4.4.2.2, "ASCII (*.csv)
on page 159
One 1-port Touchstone file per
standard and port (see
ter 4.4.2.1, "Touchstone
on page 152)
panel",
on page 606).
GUI reference
Chap-
files",
Chap-
files",
621