R&S ZNB Series User Manual page 455

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Default Port Assignment
Restores the default port assignments.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:
DEFault
Start
Proceeds to
If the configured port assignments are invalid, this action is disabled.
Step 3: Cal Unit
During the calibration phase, the R&S ZNB/ZNBT displays a "Cal Unit" screen that
guides the user through the actual correction data acquisition.
In each calibration step
the calibration unit must be connected according to the depicted port assignment;
auto-detection is possible
the related test ports are calibrated with the "Calibration Type" selected in
Ports
a subsweep is performed for each required test port or port pair and every required
standard
(and for every possible path between the port pairs, if external switch matrices are
involved)
After all these calibration steps have been completed, the resulting system error cor-
rection can be calculated and applied.
User Manual 1173.9163.02 ─ 62
Step 3: Cal
Unit.
GUI reference
Cal softtool
Step 1:
455