Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs
[SENSe<Ch>:]CORRection:COLLect:CONNection:GENDers
Import Cal Kit...
Opens the "Import Calibration Kit" dialog that allows you to import a cal kit file. For
background information, see
Remote command:
MMEMory:LOAD:CKIT
Back
Go back to
Start
Start
Step 3: Calibration
Allows you to acquire error correction data for every required port (pair) and calibration
standard, where "required" depends on the selected ports and calibration type. On
"Apply", the R&S ZNB/ZNBT calculates the system error correction (error terms) from
the measurement data of the standards and applies the result to the active channel.
The "Reduced Through" logic helps to keep the number of transmission measurements
as small as possible (see
ber of Through
User Manual 1173.9163.02 ─ 62
Step 1: Ports and
Step 3:
Calibration.
Chapter 4.5.1.11, "Full n-Port calibration with reduced num-
connections",
Chapter 4.5.2.4, "Cal kit
Type.
on page 173).
GUI reference
Cal softtool
Files",
on page 179.