R&S
ZNB/ZNBT
Common controls
Reset
Resets the "Freq. Conversion" and "Power Conversion" settings of all ports. After
"Reset", an "Apply" or "OK" terminates the arbitrary mode.
Remote command:
[SENSe<Ch>:]FREQuency:CONVersion
Edit Stimulus
For non-segmented sweeps, the "Edit Stimulus..." button opens the
that allows you to access the stimulus settings without having to leave the "Port Set-
tings" dialog.
Apply / OK / Cancel
Modified "Port Settings" take effect on "Apply" or "OK". Use "Cancel" to discard possi-
ble changes.
If R&S ZNB/ZNBT-K4 is available, then, depending on the current port configuration,
"Apply" and "OK" also deactivate the (arbitrary) frequency conversion mode.
If individual port frequencies are selected, the arbitrary mode is activated.
If individual port powers but no individual frequencies are selected, the current
mode is maintained.
If all port frequencies are equal to the channel base frequency, and all port powers
are equal to the channel base power (e.g. after Reset, the arbitrary mode is deacti-
vated.
Remote command:
[SENSe<Ch>:]FREQuency:CONVersion ARBitrary | FUNDamental
Stimulus dialog
The "Stimulus" dialog gives access to the parameters of a non-segmented sweep.
Access:
Figure 5-45: Stimulus Dialog: Linear Frequency Sweep
User Manual 1173.9163.02 ─ 62
Port Settings dialog
> "Edit Stimulus..."
Channel Config softtool
FUNDamental
GUI reference
Stimulus dialog
569