CALCulate<Chn>:LIMit:SOUNd[:STATe] <Boolean>
Deprecated command: Since version 3.45, generating sounds is no longer supported
by the analyzer firmware.
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded limit.
Suffix:
<Chn>
Parameters:
<Boolean>
Example:
CALCulate<Chn>:LIMit:STATe <Boolean>
Switches the limit check (including upper and lower limits) on or off.
Tip: Use CALCulate<Ch>:LIMit:UPPer:STATe or
CALCulate<Ch>:LIMit:LOWer:STATe to switch on or off the individual limit checks
for upper or lower limit lines.
Suffix:
<Chn>
User Manual 1173.9163.02 ─ 62
.
Channel number used to identify the active trace
Segment number
LMIN | LMAX | OFF
Limit line type
Range:
LMAX (upper limit line segment), LMIN (lower limit
line segment), OFF (limit check switched off, limit
line segment not deleted)
*RST:
LMAX
*RST; :CALC:LIM:UPP 0, 0
Define an upper limit line segment across the entire sweep
range, using a constant upper limit of 0 dBm.
CALC:LIM:SEGM:TYPE LMIN
Turn the defined limit line segment into a lower limit line seg-
ment.
See
"Segment List"
on page 378
.
Channel number used to identify the active trace
ON | OFF - Fail beep on or off.
*RST:
OFF
CALC:LIM:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
.
Channel number used to identify the active trace
Command reference
SCPI command reference
938