R&S ZNB Series User Manual page 709

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Skew
Enables/disables the skew measurement between the active trace and the
Trace
The skew measurement can only be performed if the following conditions are met for
both the active trace and the reference trace:
the trace format is
Time Domain
the "Low Pass Step" time domain transformation
stimulus grid to be harmonic)
If these conditions are not met, a message is displayed:
Instead of performing the required configuration changes manually, you may also use
the "Auto Setup" button of the corresponding info field to let the firmware do this for
you.
Remote command:
CALCulate<Chn>:DTIMe:STATe
Reference Trace
Selects the reference trace for the skew measurement.
Note that the reference trace must be in the same channel as the active trace.
See
Remote command:
CALCulate<Chn>:DTIMe:TARGet
Skew Position
Defines the percentage of the step size that defines the position of the reference points
on the current and the reference trace. The "Skew" value is the delta between the X
coordinates of these reference points.
User Manual 1173.9163.02 ─ 62
Real
is enabled
Skew
for additional conditions on both the active and the reference trace.
GUI reference
Applic softtool
Reference
Type
is used (which requires the
709