Calculate:fmodel:delt:m1L:measure; Calculate:fmodel:delt:m2L:measure; Calculate:fmodel:delt:m3L:measure - R&S ZNB Series User Manual

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Options:
Manual operation:

CALCulate:FMODel:DELT:M1L:MEASure

CALCulate:FMODel:DELT:M2L:MEASure

CALCulate:FMODel:DELT:M3L:MEASure

Measures the full set of S-parameters at the active ports (see
Del:DELT<Ph_pt>:M1L|M2L|M3L[:STATe]).
Usage:
Options:
Manual operation:
CALCulate:FMODel:DELT<Ph_pt>:M1L[:STATe] <Boolean>
CALCulate:FMODel:DELT<Ph_pt>:M2L[:STATe] <Boolean>
CALCulate:FMODel:DELT<Ph_pt>:M3L[:STATe] <Boolean>
Defines the active ports for the 1st/2nd/3rd line Delta-L measurement
late:FMODel:DELT<Ph_pt>:M1L|M2L|M3L:MEASure).
For each measurement, either two or four ports must be active. If the 1-length method
is used
the first line.
Suffix:
<Ph_pt>
Parameters:
<Boolean>
Options:
Manual operation:
CALCulate:FMODel:DELT:MEASurement <Measurement>
Defines whether single-ended or balanced ports are measured for Delta-L characteri-
zations.
Parameters:
<Measurement>
User Manual 1173.9163.02 ─ 62
R&S ZNB/ZNBT-K231
See
"Length"
Event
R&S ZNB/ZNBT-K231
See
"Measure"
(CALCulate:FMODel:DELT:METHod
.
Physical port number
ON (1)
Port <Ph_pt> is active in the respective measurement
OFF (0)
Port <Ph_pt> is inactive in the respective measurement
R&S ZNB/ZNBT-K231
See
"Ports table"
SINGleended | BALanced
SINGleended
Single-ended ports only
BALanced
Balanced ports only
on page 666
on page 665
M1L), then four ports must be active for
on page 665
Command reference
SCPI command reference
CALCulate:FMO
(CALCu
899

Advertisement

Table of Contents
loading