This calibration type requires a Short, an Open and a Match standard to be connected
to a single test port plus a Through standard between this calibrated source port and
the other load ports. The four standard measurements are used to derive the following
error terms:
The Short and Open standards are used to derive the source match and the reflec-
tion tracking error terms at the source port.
The Match standard is used to derive the directivity error at the source port.
The Through standard provides the transmission tracking error terms.
For calibration of two ports, a one-path two-port calibration requires only four standards
to be connected (instead of 7 for a full two-port TOSM calibration). It is suitable when
only the forward (e.g. S
ded, and if the DUT is well matched, especially at the load port.
A one-path two-port calibration is also the best calibration method for test setups with
unidirectional signal flow.
4.5.1.4

TOSM and UOSM calibration

TOSM
A TOSM (Through – Open – Short – Match) calibration requires the same standards as
the one path two ports calibration, however, all measurements are performed in the for-
ward and reverse direction. TOSM is also referred to as SOLT (Short – Open – Load =
Match – Through) calibration. The four standards are used to derive 6 error terms for
each signal direction:
In addition to the source match and reflection tracking error terms provided by the
one-path two-port calibration, TOSM also provides the load match.
The directivity error is determined at all source ports.
The transmission tracking is determined for each direction.
TOSM calibration is provided for arbitrary n-port measurements (n>1). The number of
required standard measurements and of error terms increases as shown in the follow-
ing table.
Number of ports
2
3
4
*) No isolation terms are available.
An Open, Through and Match measurement is required at each port; in addition, for a
full TOSM, a Through must be measured for every directed port pair. Therefore the
User Manual 1173.9163.02 ─ 62
and S
) or reverse S-parameters (e.g. S
11
21
Number of standards
Number of standard
to be connected
measurements
2 * 3
2 * 3
+1 = 7
+2 * 1 = 8
3 * 3
3 * 3
+ 2 + 1 = 12
+2 * (2 + 1) = 15
4 * 3
4 * 3
+3 + 2 + 1 = 18
+2 * (3 + 2 + 1) = 24
Concepts and features
Calibration
and S
) are nee-
22
12
Number of error terms
2 * 3
+ 2 * 2 = 10
3 * 3
+ 2 * 2 * 3 = 21
4 * 3
+ 2 * 2 * 6 = 36
167
*)