CALCulate<Chn>:LIMit:DCIRcle:DATA <CenterX>, <CenterY>, <Radius>
Defines the display circle for the active trace of channel <Chn>.
Suffix:
<Chn>
Parameters:
<CenterX>
<CenterY>
<Radius>
Manual operation:
CALCulate<Chn>:LIMit:DCIRcle:DISPlay[:STATe] <Boolean>
Sets/queries the visibility of the display circle for the active trace of channel <Chn>.
The display circle is defined using CALCulate<Chn>:LIMit:DCIRcle:DATA.
Suffix:
<Chn>
Parameters:
<Boolean>
Manual operation:
CALCulate<Chn>:LIMit:DCIRcle[:STATe] <Boolean>
Sets/queries the state of the display circle for the active trace of channel <Chn>.
The display circle is defined using CALCulate<Chn>:LIMit:DCIRcle:DATA.
Suffix:
<Chn>
Parameters:
<Boolean>
Manual operation:
CALCulate<Chn>:LIMit:DELete:ALL
Deletes all limit line segments.
Suffix:
<Chn>
User Manual 1173.9163.02 ─ 62
.
Channel number used to identify the active trace
X position (real part) of the display circle's center
Y position (imaginary part) of the display circle's center
Radius of the display circle
See
"Draw Circle / Radius, Center X, Center Y"
.
Channel number used to identify the active trace
If set to ON, the line of the display circle is shown whenever the
related trace is displayed in complex format.
See
"Show Border"
on page 389
.
Channel number used to identify the active trace
If set to ON, only trace points within the display circle are shown
at the GUI whenever the related trace is displayed in complex
format.
See
"Limit to Circle On/Off"
.
Channel number used to identify the active trace
Command reference
SCPI command reference
on page 390
on page 389