Calibrate Only Port Frequency
For
Frequency conversion
brated for all resulting frequencies – even if they are currently not relevant for some
ports. Using this switch, you can limit the calibrated frequencies to those frequencies
that are required for the related port, which results in shorter calibration times.
Power Cal dialog – Meas. Receiver
In "Meas. Receiver" mode, the "Power Cal" dialog guides you through the calibration of
a measurement receiver. To ensure accurate source signal powers, a source power
calibration must be performed before the measurement receiver calibration.
Background information
Refer to
Access:
left: = source and meas. port identical
right = source and meas. port different
Port Overview
The dialog shows all receiver ports of the network analyzer. The selected port is dis-
played with the current cal power settings (see
dialog",
surement receiver calibration.
User Manual 1173.9163.02 ─ 62
measurements, by default the reference receivers are cali-
Chapter 4.5.6.2, "Measurement receiver
Power Cal wizard
> "Meas. Receiver"
on page 527); moreover, a circuit diagram visualizes the purpose of the mea-
calibration",
on page 196.
Chapter 5.11.3.2, "Modify Cal Power
GUI reference
Cal softtool