The table defines all settings related to "Multiple Triggers" (see
ger
tab",
are available:
"Free Run" means that a trigger signal is not required
"External 1" is the external trigger signal fed in at either the Trigger In connector on
the rear panel or pin 2 of the User Port connector.
"External 2" is the external trigger signal fed in at pin 25 of the User Port connector
on the rear panel.
"External 1 and 2"/"External 1 or 2" means that the measurement sequence is initi-
ated after the analyzer has received an event from external trigger signal 1 and/or
2. A
Manual means that the trigger event is generated manually, by selecting the
ual Trigger
For example, a new sweep can be triggered by an external trigger no. 1, while the indi-
vidual sweep points are triggered by external trigger no. 2. External trigger 1 is ignored
if a sweep is running, external trigger 2 is ignored if there is no running sweep. In the
figure below a sweep comprises 5 measurement points and dotted arrows depict
ignored trigger events.
Remote command:
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce
TRIGger<Ch>[:SEQuence]:MULTiple:SLOPe<Num>
TRIGger<Ch>[:SEQuence]:MULTiple:HOLDoff
5.10.4 Sweep Control tab
Allows you to control the sweep execution.
The availability of certain settings depends on the selected
User Manual 1173.9163.02 ─ 62
on page 437). For all measurement sequences the following trigger sources
Signal Type
must be specified for both trigger signals.
button.
GUI reference
Sweep Softtool
Chapter 5.10.3, "Trig-
Man-
Remote
Language.