R&S ZNB Series User Manual page 314

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Power Meter
Shows a list of all power meters that have been properly configured. See
Devices"
The bordered label below displays the type and serial number of the selected power
meter.
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:SDEFine
Auto Zero
Initiates an automatic zeroing procedure of the selected power meter.
The power meter must be disconnected from the RF power; see
"Zeroing",
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
Source Port
Selects one of the available test ports of the analyzer as a source of the stimulus sig-
nal.
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:SDEFine
Show as
Selects the physical unit of the displayed trace. It is possible to display the measured
"Voltage" V or convert it to a "Power" according to the formula
P = V
Z
denotes the reference impedance of the source port. The reference impedances are
0
defined in the "Balanced Ports" dialog (see
on page 289).
Remote command:
CALCulate<Chn>:FORMat:WQUType
5.2.11 DC tab
Selects the ranges for the DC voltages fed to the DC Input connectors at the rear panel
as measured quantities and opens the PAE measurement dialog. The DC input con-
nectors are available as option R&S ZNB/ZNBT-B81 (see
on page 262).
User Manual 1173.9163.02 ─ 62
on page 782.
on page 266. A message indicates that zeroing is finished.
2
/Re(Z
).
0
'TraceName', 'PmtrD1 | ...
'TraceName', 'PmtrD1 | ...
Chapter 4.7.27.1,
'TraceName', 'PmtrD1 | ...
'TraceName', 'PmtrD1 | ...
Chapter 5.2.1.3, "Balanced Ports
Chapter 4.7.25, "DC
GUI reference
Meas softtool
"Configured
dialog",
inputs",
314