R&S ZNB Series User Manual page 537

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Ch<n> Calibration Properties/Cal Group <n> Properties
Displays the basic channel settings and the properties of the system error correction
and the power correction for the channel (calibration group) selected in the "Channel
State" ("Pool") table.
"Ch<n> Calibration Properties" also indicates settings mismatches (between the
selected channel and the channel setup that was used during calibration) that lead to a
"Cal Off". Currently this indication is limited to settings related to Parallel Measure-
ments with
In addition, it is stated whether sweep data are available for the selected calibration.
Remote command:
[SENSe<Ch>:]CORRection:DATE?
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>?
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>:PORT<PhyPt>?
[SENSe<Ch>:]CORRection:DATA:PARameter:COUNt?
[SENSe<Chn>:]CORRection:SSTate?
5.11.4.4
JoinCal dialog
Allows you to join existing calibrations.
Access: Channel – [Cal] > "Use Cal" > "JoinCal"
Background information
See
Input Calibration1/Input Calibration2/Cal Name
Select the input calibrations to be joined, and specify a name for the resulting calibra-
tion
Remote command:
SYSTem:CORRection:JOINcal
User Manual 1173.9163.02 ─ 62
Frequency Offset
Chapter 4.5.9, "Joining
and
Freq Sweep
Mode.
calibrations",
on page 202.
GUI reference
Cal softtool
537

Advertisement

Table of Contents
loading

Table of Contents