R&S
ZNB/ZNBT
Factor / On / Reset
"Factor" defines the number of averaged traces, "On" enables or disables the sweep
average, "Reset" starts a new average cycle. The average cycle is also restarted when
the
averaging mode
Remote command:
[SENSe<Ch>:]AVERage:COUNt
[SENSe<Ch>:]AVERage[:STATe]
[SENSe<Ch>:]AVERage:CLEar
Mode
Selects one of the following averaging algorithms:
"Auto": Automatic selection between "Reduce Noise" and "Flatten Noise"
mode, depending on the trace type.
"Reduce Noise": Cumulative moving averages of the real and imaginary parts of
each measurement result, provides the most effective noise suppression for the
"Real" and "Imag" formats and for polar diagrams.
"Flatten Noise": Cumulative moving averages of the (linear) magnitude and phase
values, provides the most effective noise suppression for the "dB Mag", "Phase",
"Unwr. Phase", and "Lin Mag" formats.
"Moving Average": Simple moving averages of the real and imaginary parts of
each measurement result; similar to "Reduce Noise", but with finite history.
Changing the mode resets the average cycle.
Note: For frequency conversion measurements, always "Flatten Noise" is used.
Remote command:
[SENSe<Ch>:]AVERage:MODE
5.10 Sweep Softtool
The "Sweep" softtool allows you to define the scope of the measurement in the active
channel.
User Manual 1173.9163.02 ─ 62
is changed.
GUI reference
Sweep Softtool
422