You can also use the basic transmission test setup, e.g. if you want to measure
reflection and transmission parameters in parallel.
The analyzer provides special calibration types for reflection measurements. Use
the calibration wizard and select an appropriate type.
A full n-port calibration (TOSM, UOSM, TNA ...) corrects the system errors for all
transmission and reflection S-parameters.
Some of the trace formats are particularly suited for reflection measurements. For
instance, you can display the measured reflection coefficient S
obtain the complex input impedance at port 1.
Proceed as described in
on page 28 to shut down your analyzer.
User Manual 1173.9163.02 ─ 62
VNA
1
DUT
1
Chapter 3.1.7, "Switching the instrument on and
Getting started
Performing measurements
in a Smith chart to
11
off",
86