Calculate:fmodel:delt:sweep:control; Calculate:fmodel:delt:sweep:control:ifbw; Calculate:fmodel:delt:sweep:control:step - R&S ZNB Series User Manual

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
The resulting frequency step size and IF bandwidth can be queried using
FMODel:DELT:SWEep:CONTrol:STEP?
DELT:SWEep:CONTrol:IFBW?
Parameters:
<SweepControl>
Options:
Manual operation:

CALCulate:FMODel:DELT:SWEep:CONTrol:IFBW?

For the Delta-L measurement, the firmware can either use a default IF bandwidth, or
the active channel setting (see CALCulate:FMODel:DELT:SWEep:CONTrol).
This query returns the resulting IF bandwidth.
Usage:
Options:
Manual operation:

CALCulate:FMODel:DELT:SWEep:CONTrol:STEP?

For the Delta-L measurement, the firmware can either use a default frequency step
size, or the active channel setting (see
).
This query returns the resulting frequency step size.
Usage:
Options:
Manual operation:
CALCulate:FMODel:DELT:SWEep:FREQuency:MAXimum <Frequency>
CALCulate:FMODel:DELT:SWEep:FREQuency:MINimum <Frequency>
Defines the sweep range of the Delta-L measurement channel. The entered MAXimum
and MINimum values are set as
[SENSe<Ch>:]FREQuency:STARt, respectively.
Furthermore, the VNA collects data for the LPI computation a the MAXimum and MINi-
mum frequencies (in addition to
CURRent?).
User Manual 1173.9163.02 ─ 62
on page 901 and
on page 901, respectively.
DEFault | CURRent
DEFault
Use the default values of the Delta-L algorithm:
10 MHz frequency step size, 1 kHz IF bandwidth
SweepControl
Use the values of the active channel
R&S ZNB/ZNBT-K231
See
"Sweep Control"
on page 663
Query only
R&S ZNB/ZNBT-K231
See
"Sweep Control"
on page 663

CALCulate:FMODel:DELT:SWEep:CONTrol

Query only
R&S ZNB/ZNBT-K231
See
"Sweep Control"
on page 663
[SENSe<Ch>:]FREQuency:STOP
CALCulate:FMODel:DELT:FREQuencies:
Command reference
SCPI command reference
CALCulate:
CALCulate:FMODel:
and
901

Advertisement

Table of Contents
loading

Table of Contents