R&S ZNB Series User Manual page 886

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
CALCulate<Chn>:FILTer[:GATE]:TIME:STATe <Boolean>
Determines whether the time gate for trace no. <Chn> is enabled.
Suffix:
<Chn>
Parameters:
<Boolean>
Example:
Manual operation:
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE] <TimeGateFilter>
Selects the time gate filter type, defining what occurs to the data in the specific time
region.
Suffix:
<Chn>
Parameters:
<TimeGateFilter>
Example:
Manual operation:
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow <TimeGate>
Selects the time gate to be applied to the time domain transform.
User Manual 1173.9163.02 ─ 62
.
Channel number used to identify the active trace
ON - time gate enabled
OFF - time gate disabled
*RST:
OFF
*RST; :CALC:TRAN:TIME:STAT?
CALC:FILT:TIME:STAT?
Reset the instrument, activating a frequency sweep, and query
whether the default trace is displayed in the time domain and
whether the time gate is enabled. The response to both queries
is 0.
See
"Time Gate"
on page 353
.
Channel number used to identify the active trace
BPASs | NOTCh
BPASs
Band pass filter: Pass all information in specified time region and
reject everything else
NOTCh
Notch filter: Reject all information in specified time region and
pass everything else
*RST:
BPASs
*RST; :CALC:FILT:TIME:STAT ON
Reset the instrument and enable the time gate.
CALC:FILT:TIME NOTCh
Select a notch filter in order to reject unwanted pulses.
See
"Bandpass / Notch"
Command reference
SCPI command reference
on page 354
886

Advertisement

Table of Contents
loading

Table of Contents