displays the corresponding error code and description (see
]SWEep:GENeration:ANALog:CONDition?)
It is essential to perform the calibration with the same "Freq Sweep Mode" settings
as for the actual measurement; otherwise the calibration is deactivated ("Cal Off").
The
mismatches.
A scalar power calibration cannot be created or used in swept mode.
Remote command:
[SENSe<Ch>:]SWEep:GENeration
[SENSe<Ch>:]SWEep:GENeration:ANALog:CONDition?
5.10.2 Sweep Type tab
Defines the sweep variable (frequency/power/time) and the position of the sweep
points across the sweep range.
5.10.2.1
Controls on the Sweep Type tab
Lin Freq
In a linear frequency sweep, the stimulus frequency is swept in equidistant steps over
the continuous frequency range. The frequency range (sweep range) and the internal
generator power can be specified in the "Stimulus" settings (see
lus
tab",
stant and given by <Span>/(n - 1) where n is the specified "Number of Points" (n > 1).
"Lin Freq" is the default sweep type. In a Cartesian diagram, the measurement result is
displayed as a trace over a linear frequency scale (as known, e.g., from spectrum ana-
lyzers). The following example shows a "Lin Freq" sweep with the forward transmission
parameter S
User Manual 1173.9163.02 ─ 62
Calibration Manager dialog
on page 414). The step width between two consecutive sweep points is con-
as measured quantity, and a "dB Mag" scaled y-axis.
21
provides concise information on possible mode
GUI reference
Sweep Softtool
[SENSe<Ch>:
Chapter 5.8.1, "Stimu-
426