R&S ZNB Series User Manual page 505

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
If inactive, the R&S ZNB/ZNBT permanently refreshes all traces of all diagrams
which can put a heavy load on the connected switch matrices (if any).
Hence it is recommended (and default) to activate it, in particular if one of the matrices
uses mechanical switches (that wear off apart from making noise).
Apply
Is enabled when sufficient data have been acquired for the calibrated ports and stand-
ards and for the power meter. The button starts the calculation of the calibration data
and closes the calibration wizard. The current instrument settings are stored with the
correction data.
To avoid incompatibilities, older calibration data is deleted unless it has been transfer-
red into the "Cal Pool" using the "Calibration Manager".
Use the
Chapter 4.5.7.3, "Combining SMARTerCal with scalar power
Note: Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel
settings can cause the calibration to be inaccurate. The analyzer auto-detects potential
sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:DUMMy]
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault
[SENSe<Ch>:]CORRection:COLLect:DELete
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>?
[SENSe<Ch>:]CORRection:DATA:PARameter<Sfk>:PORT<PhyPt>?
[SENSe<Ch>:]CORRection:DATE?
[SENSe<Chn>:]CORRection:SSTate?
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
5.11.1.11
SMARTerCal wizard (manual, MultiCal)
The multiple calibration version of the
through the setup and simultaneous manual acquisition of multiple SMARTerCals.
Except for the possibility to create multiple calibrations, the layout and functionality is
the same as for the single calibration version of the wizard.
Access: Channel – [Cal] > "Start Cal" > "SMARTerCal" – "Start... (Manual)"
Background and related information
The "Multiple Cal in Calibration Wizard" feature is disabled by default and has to be
activated in the
See
information.
User Manual 1173.9163.02 ─ 62
Power Cal wizard
to perform an additional source flatness calibration (see
Calibration tab
Chapter 5.11.1.10, "SMARTerCal wizard
SMARTerCal wizard (manual)
of the "System Config" dialog.
(manual)",
GUI reference
Cal softtool
calibration",
on page 200).
guides you
on page 496 for more
505