R&S ZNB Series User Manual page 984

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Parameters:
<StimValue>
Example:
Manual operation:
CALCulate<Chn>:RIPPle:SOUNd[:STATe] <Boolean>
Deprecated command: Since version 3.45, generating sounds is no longer supported
by the analyzer firmware.
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded ripple limit.
Suffix:
<Chn>
Parameters:
<Boolean>
Example:
CALCulate<Chn>:RIPPle:STATe <Boolean>
Switches the ripple limit check for the active trace on or off.
Tip: Use
check for a single ripple limit range on or off.
Suffix:
<Chn>
Parameters:
<Boolean>
Example:
User Manual 1173.9163.02 ─ 62
Stimulus values (unitless) confining the ripple limit range.
The unit is adjusted to the sweep type of the active channel
([SENSe<Ch>:]SWEep:TYPE).
Range:
*RST:
Default unit: NN
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:SEGM:STIM:STAR 1GHZ; STOP 2.5 GHZ; :
CALC:RIPP:SEGM:LIM 5
Change the range to a stimulus range between 1 GHz and 2.5
GHz and a limit of 5 dB.
CALC:RIPP:SEGM:STAT OFF
Disable the limit check in the modified stimulus range.
See
"Range List"
.
Channel number used to identify the active trace
ON | OFF - fail beep on or off.
*RST:
CALC:RIPP:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe]
.
Channel number used to identify the active trace
ON | OFF – ripple limit check on or off.
*RST:
See
CALCulate<Chn>:RIPPle:FAIL?
Virtually no restriction for ripple limit ranges.
n/a (no ripple limit line defined after a reset)
on page 384
OFF
OFF
Command reference
SCPI command reference
to switch the limit
984