R&S ZNB Series User Manual page 883

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Example:
Manual operation:
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer <CenterTime>
Defines the center time of the time gate.
Suffix:
<Chn>
Parameters:
<CenterTime>
Example:
Manual operation:
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev <SidebandSupp>
Sets the sideband suppression for the Dolph-Chebyshev time gate. The command is
only available if a Dolph-Chebyshev time gate is active
FILTer[:GATE]:TIME:WINDow
Suffix:
<Chn>
User Manual 1173.9163.02 ─ 62
*RST; :CALCulate1:TRANsform:TIME:STATe ON
CALCulate1:FILTer:GATE:TIME:STATe ON; SHOW ON
Activate time domain representation and a time gate in channel
no. 1. Display the time gate
CALCulate1:FILTer:GATE:TIME:STARt 2ns; STOP 3
ns
Restrict the time gate to the time interval between 2 ns and 3 ns.
CALCulate:FILTer:GATE:TIME:AOFFset ON
Activate an offset of the time gate according to a new delay set-
ting.
SENSe1:CORRection:EDELay1:TIME 1ns
Specify a 1 ns delay at port 1.
CALCulate1:FILTer:GATE:TIME:STARt?; STOP?
Query the time gate position. The response is 1E-009;2E-009.
See
"Adjust Time Gate"
.
Channel number used to identify the active trace
Center time of the time gate
Range:
-99.8999999 s to +99.8999999 s
Increment: 0.1 ns
*RST:
1.5E-009 s
Default unit: s
*RST; :CALC:TRAN:TIME:STAT ON; :CALC:FILT:TIME:
STAT ON
Reset the instrument and enable the time domain representation
and the time gate.
CALC:FILT:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.
See
"Axis Pair"
on page 354
DCHebyshev).
.
Channel number used to identify the active trace
Command reference
SCPI command reference
on page 617
(CALCulate<Chn>:
883

Advertisement

Table of Contents
loading

Table of Contents