"PASS": all measured reflection parameters are within the specified
its, compared to the selected reference characterization.
"FAIL" otherwise.
[In this case, the validation stops at the first failed reflection measurement.]
Remote command:
CALCulate:CALValidate:RUN
Advanced Mode
Switches to
Advanced mode
In advanced mode you can configure the validation settings (Characterization,
tion
Limits) and run or rerun particular measurements (Measurement,
Standard"
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Measurement
Selects the measurement to be performed in advanced mode and runs it
Characterization
Selects the calibration unit characterization to be used for the validation.
By default the factory characterization is used.
Remote command:
CALCulate:CALValidate:CHARacteriza
Calibration Standard
Selects the calibration standard to be measured in advanced mode.
Validation Limits
The "Limits..." button opens the "Cal Validation Limits" dialog that allows you to specify
the maximum tolerable deviations from the corresponding S-parameters of the
selected reference characterization.
User Manual 1173.9163.02 ─ 62
Advanced
mode.
on page 510 ).
GUI reference
Cal softtool
Validation Lim-
Valida-
"Calibration
510