Opens a dialog that allows to set the following parameters:
"Wait for Data after Sweep" determines the execution behavior of
INITiate[:IMMediate] commands (see
on page 1132).
If enabled, an automatic *WAI is added (see
on page 842).
By default this is disabled.
"Auto-Align Logical Ports" determines the logical port creation logic.
If enabled (default), logical ports are aligned and must be set from low to high port
(which was the only possibility prior to firmware V1.91). If set to disabled, new ports
can be created freely, like in manual operation.
"No Device Clear" suppresses Device Clear (DCL, SDC) GPIB interface mes-
sages.
Remote command:
SYSTem:COMMunicate:GPIB[:SELF]:INIT:WAIT
SYSTem:COMMunicate:GPIB[:SELF]:LPORt:ALIGn
SYSTem:COMMunicate:GPIB[:SELF]:DCLear:SUPPress
Instrument Messages...
Opens the System Configuration dialog with the
you can configure the display of instrument messages as information popups, in partic-
ular the display of remote control command errors.
5.18.4 External Devices tab
Allows you to set up and configure external power meters, generators and calibration
units and to enable error logging for the current session.
Background information
Refer to the following sections:
Chapter 4.7.27, "External power
Chapter 4.7.28, "External
Chapter 4.5.5, "Automatic
User Manual 1173.9163.02 ─ 62
Chapter 7.3.8, "INITiate
Chapter 7.2, "Common