The "Export Eye Measurements – ASCII Files" dialog is a standard "Save File" dialog
with the additional options of selecting a "Field Separator" (semicolon, comma, tab or
space) and a "Decimal Separator" (comma or point).
Note that the decimal separator and field separator must be different.
Remote command:
MMEMory:STORe:EYE:MEASurements
5.16.2.5
Eye Diagram dialog
Allows basic configuration of the eye diagram simulation.
Access: System – [Applic] > "TDR" > "Eye Diag" > "Eye Diagram..."
For advanced configuration of the eye diagram simulation, see
"Advanced Settings
Measurement / Topology
"Measurement" allows you to select an S-Parameter, i.e. the transmission (or reflec-
tion) whose time domain properties you want to analyze. The "Topology" to the right is
updated accordingly.
Remote command:
CALCulate<Ch>:PARameter:MEASure
Bit Stream
Describes the bit stream generated by the "virtual" generator.
"Bit Stream" can either be a pseudo-random binary sequence (PRBS) of the selected
Length
The PRBSs are generated using a linear feedback shift register with generator polyno-
mials allowing the maximal run length of the sequence.
User Manual 1173.9163.02 ─ 62
dialog",
on page 695
or a user-defined bit stream.
GUI reference
Applic softtool
Chapter 5.16.2.6,
692