R&S ZNB Series User Manual page 660

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
The ports used for the measurement must be unique. That is, you cannot select P1
for both sides of a "Sym 2x Thru" coupon. However, you can select the same port
for both measurements in a "1x Open, 1x Short" coupon, which requires a connec-
tion change before the second measurement.
The "Load" button allows you to load a Touchstone file instead of measuring the
coupon. The graphic displays the Touchstone file's required "cardinality".
Remote command:
CALCulate:FMODel:DEASsistant:COUPon:LEFT:TYPE
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:TYPE
CALCulate:FMODel:DEASsistant:COUPon:LEFT:PORT
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:PORT
CALCulate:FMODel:DEASsistant:COUPon:LEFT:MEASure
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:MEASure
CALCulate:FMODel:DEASsistant:COUPon:LEFT:FILE
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:FILE
CALCulate:FMODel:DEASsistant:COUPon:LEFT:CLEar
CALCulate:FMODel:DEASsistant:COUPon:RIGHt:CLEar
Measure / Apply section
After you have measured the coupons (or loaded their data from file), you can proceed
with the DUT. Choose:
A single-ended or balanced port for the 1x DUT sides
Two single-ended ports for the 2x single-ended DUT sides
A range of single-ended or balanced ports for the nx DUT sides
If you want to apply impedance correction, select "Impedance Correction" and "Mea-
sure" the DUT.
Once the coupons (and the DUT) are measured, the "Apply" button becomes active.
Use it to run the tool and, if successful, perform the deembedding on the selected
ports.
User Manual 1173.9163.02 ─ 62
For single-ended coupons, any physical port can be selected, even if that port
is part of a logical balanced port or port set.
For balanced coupons, you can only select a preciously defined logical port.
GUI reference
Offset Embed softtool
660

Advertisement

Table of Contents
loading

Table of Contents