R&S ZNB Series User Manual page 998

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Suffix:
<Chn>
Parameters:
<PeakThreshold>
Example:
Options:
Manual operation:
CALCulate<Chn>:TRANsform:DTFault:POINts
If the active trace of channel <Chn> is a Distance to Fault (DtF) trace, this command
starts the "Auto Number of Points" calculation (see
on page 729).
An execution error is raised, if the calculated number of points is higher than the maxi-
mum number of sweep points the firmware allows (100001). To avoid this, reduce the
frequency span Δf (see
stop frequency d
2.6 · d
where v denotes the velocity factor of the measured cable (see
TRANsform:DTFault:SELect).
Suffix:
<Chn>
Example:
Usage:
Options:
Manual operation:
User Manual 1173.9163.02 ─ 62
.
Channel number used to identify the active trace
Fault limit
*RST:
Default unit: dB
*RST; :CALCulate1:TRANsform:DTFault:STATe ON
makes the active trace of channel 1 a DtF trace.
CALCulate1:TRANsform:DTFault:PEAK:THReshold -30
sets the DtF limit to -30 dB.
R&S ZNB/ZNBT-K3
See
"Fault Limit"
[SENSe<Ch>:]SWEep:POINts
(see CALCulate<Chn>:TRANsform:DTFault:STOP) so that
stop
· Δf / (v · c
) ≤ 100001 ,
stop
0
.
Channel number used to identify the active trace
*RST; :CALCulate1:TRANsform:DTFault:STATe ON
makes the active trace of channel 1 a DtF trace.
SENSe1:SWEep:POINts?
returns the default number of sweep points (201).
:CALCulate1:TRANsform:DTFault:STARt 0; STOP 100
widens the DtF distance window to 100m.
CALCulate1:TRANsform:DTFault:POINts
starts the "Auto Number of Points calculation". Now
SENSe1:SWEep:POINts?
returns an increased number of sweep points.
Event
R&S ZNB/ZNBT-K3
See
"Auto Number of Points"
-20 dB
on page 732
"Auto Number of Points"
on page 1324) and/or the
on page 729
Command reference
SCPI command reference
CALCulate<Chn>: