R&S ZNB Series User Manual page 995

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
See also
TRANsform:DTFault:STOP.
Suffix:
<Chn>
Parameters:
<Center>/<Span>
Example:
Options:
Manual operation:
CALCulate<Chn>:TRANsform:DTFault:DEFine <CblName>, <CblPermittivity>,
Defines a new user-defined cable type for Distance to Fault (DtF) measurements.
The cable's frequency-dependent attenuation is defined via attenuation/frequency pairs
<CblAtt1>, <CblFreq1>, <CblAtt2>, <CblFreq2>, ... . At least one pair
has to be specified.
Suffix:
<Chn>
Setting parameters:
<CblName>
<CblPermittivity>
<CblAtt[i]>
<CblFreq[i]>
Example:
Usage:
User Manual 1173.9163.02 ─ 62
CALCulate<Chn>:TRANsform:DTFault:STARt
.
Channel number used to identify the active trace.
Center/span value.
Default unit: m
:CALCulate1:TRANsform:DTFault:CENTer 2; SPAN 2
Defines the DtF distance window of the active trace of channel 1
via its center (2m) and span (2m).
R&S ZNB/ZNBT-K3
See
"Start Distance / Stop Distance"
<CblAtt1>, <CblFreq1>{, <CblAtt2>, <CblFreq2>}...
.
Channel number
This suffix is ignored: cable types are defined for all channels.
Name of the cable type.
Must be unique among all DtF cable types, with case insensitive
match.
Relative permittivity ε
Attenuation value i in the frequency-dependent attenuation table
of the cable type.
Default unit: dB/m
Frequency i in the frequency-dependent attenuation table of the
cable type.
Default unit: Hz
CALCulate:TRANsform:DTFault:DEFine 'My cable
type', 1.1, 0.01, 1GHz, 0.015, 2GHz, 0.012,
3GHz
Defines a new cable type "My cable type" with relative permittiv-
ity ε
=1.1 and the frequency-dependent attenuation given in the
r
table below.
Setting only
Command reference
SCPI command reference
and
CALCulate<Chn>:
on page 729
of the cable type.
r
995