Power tab

The settings in this tab define global power settings of the VNA.
Power Reduction at Sweep End
The power reduction settings apply to all sweep modes but are particularly useful in
single sweep mode.
Power
Figure 5-89: Power control at sweep end in Single Sweep mode (power sweep)
Note: By default, at sweep end the output power of the first measurement point is
restored ("Auto" mode). After a small settling time (~10 ms), the sweep can be restar-
ted.
Selecting a different behavior can result in a significantly longer total measurement
time, in particular if a long "Settling Delay" is used. Do not forget to revert the
Mode at Sweep End
power at the sweep end. A preset is not sufficient for this purpose.
User Manual 1173.9163.02 ─ 62
TTL1 Pass / TTL2 Pass
Keep
Auto
Reduce
End Measurement
to "Auto" mode if it is not necessary to "Reduce" or "Keep" the
for new traces.
Settling Delay
Restart Sweep
GUI reference
Setup softtool
Time
Power
761